The FDA has updated this Safety Communication following Contec’s new software patch and the firm’s Security Advisory Notice to fix the cybersecurity vulnerabilities. The patch fully removes networking functionality from the affected Contec and Epsimed devices, making them only usable for local monitoring (vital signs only observable in the physical presence of the patient).
Date Issued: January 30, 2025
The U.S. Food and Drug Administration (FDA) is raising awareness among health care providers, health care facilities, patients, and caregivers that cybersecurity vulnerabilities in Contec CMS8000 patient monitors and Epsimed MN-120 patient monitors (which are Contec CMS8000 patient monitors relabeled as MN-120) may put patients at risk after being connected to the internet.
Three cybersecurity vulnerabilities have been identified:
These cybersecurity vulnerabilities can allow unauthorized actors to bypass cybersecurity controls, gaining access to and potentially manipulating the device.
The FDA is not aware of any cybersecurity incidents, injuries, or deaths related to these cybersecurity vulnerabilities at this time.
Patient monitors are used in health care and settings for displaying information, such as the vital signs of a patient, including temperature, heartbeat, and blood pressure.
Three cybersecurity vulnerabilities have been identified, whose potential impacts fall into two main categories. A vulnerable device could be exploited to:
The vulnerabilities could allow all vulnerable Contec and Epsimed patient monitors on a given network to be exploited at the same time.
Additionally, the software on the patient monitors includes a backdoor. “Backdoor” is the term used to describe hidden functionality that device users are not told and can allow unauthorized actors to bypass cybersecurity controls. The unauthorized actors could access and potentially manipulate the device. Given the backdoor, the device and/or the network to which the device has been connected may have been or could be compromised.
Also, the FDA has authorized these patient monitors only for wired functionality (that is, ethernet connectivity). However, the FDA is aware that some patient monitors may be available with wireless (that is, WiFi or cellular) capabilities without FDA authorization.
The Cybersecurity and Infrastructure Security Agency (CISA) has identified that once the patient monitor is connected to the internet, it begins gathering and exfiltrating (withdrawing) patient data outside of the health care delivery environment, including when the device is used in a setting. The FDA and CISA continue to work with Contec to correct these vulnerabilities as soon as possible.
The unique device identifier helps identify individual medical devices, including patient monitors, sold in the United States from manufacturing through distribution to patient use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified, and problems potentially corrected more quickly.
You can identify the devices affected by checking the unique device identifier (UDI), which is a unique numeric or alphanumeric code that generally includes a device identifier (DI) that identifies the labeler and the specific version or model of a device.
The FDA takes seriously any reports of cybersecurity vulnerabilities in medical devices and will continue to work with Contec and CISA to correct these vulnerabilities as soon as possible.
The FDA will continue to assess new information concerning the vulnerabilities and will keep the public informed if significant new information becomes available.
The FDA updated this communication to announce the availability of Contec’s software patch to fix the cybersecurity vulnerabilities.
The FDA issued this communication to alert health care providers, health care facilities, patients, and caregivers of cybersecurity vulnerabilities in Contec CMS8000 patient monitors and Epsimed MN-120 patient monitors (which are Contec CMS8000 patient monitors relabeled as MN-120) that may put patients at risk after being connected to the internet.
If you think you had a problem with a Contec CMS8000 or Epsimed MN-120 patient monitors, the FDA encourages you to report the problem through the MedWatch Voluntary Reporting Form .
Health care personnel employed by facilities that are subject to the FDA's user facility reporting requirements should follow the reporting procedures established by their facilities.
If you have questions, CDRH's Division of Industry and Consumer Education (DICE) .